Tutorial: Resonator Spectroscopy#
This tutorial walks through the simplest end-to-end experiment in the toolkit: a resonator spectroscopy sweep. By the end you will have run a real calibration node, understood how the pieces fit together, and know where the results are stored.
Hardware required. This notebook talks to the Qblox cluster and a Redis database, so it is not executed when the documentation is built — the cells are shown for reference. Run it yourself on the lab PC during a booked QPU time slot, after setting your own
REDIS_PORT.
What resonator spectroscopy measures#
Each qubit is read out through a dispersive readout resonator. Sweeping the readout-line frequency across the resonator and recording the transmitted signal \(|S_{21}|\) reveals a Lorentzian dip at the resonance frequency. Fitting that dip gives us:
the resonator frequency (where we should read out),
the loaded and external quality factors \(Q_l\), \(Q_e\),
the asymmetry phase \(\theta\).
These are the first things you calibrate on a fresh chip: without a good readout frequency, nothing else can be measured reliably. The fit model is a Lorentzian dip on a linearly tilted background, so traces with sloped shoulders are handled correctly (see bragi_quantum.experiments.nodes.readout.resonator_spectroscopy.analysis).
1. Imports#
The experiment is assembled from three packages:
bragi_quantum.api provides the
CalibrationOrchestrator(the driver) and theRedisDataHandler(where data is persisted).bragi_quantum.datalayer provides
SessionOptions, which configures a single run.bragi_quantum.experiments provides the
ResonatorSpectroscopynode together with itsParametersandQOIclasses.
[ ]:
from bragi_quantum.api.handlers.data import RedisDataHandler
from bragi_quantum.api.orchestra import CalibrationOrchestrator
from bragi_quantum.datalayer.managing.session import SessionOptions
from bragi_quantum.experiments.nodes.readout.resonator_spectroscopy.node import (
ResonatorSpectroscopy,
)
from bragi_quantum.experiments.nodes.readout.resonator_spectroscopy.parameters import (
ResonatorSpectroscopyParameters,
ResonatorSpectroscopyQOI,
)
2. Configuration#
Pick the qubits to measure and set your own reserved Redis port so you do not write into someone else’s database. HALF_WIDTH is the half-width of the frequency window swept around each qubit’s known readout frequency.
[ ]:
QUBITS = [f"q{i:02d}" for i in range(16, 26)] # q16 ... q25
SAMPLES = 101 # frequency points per resonator
HALF_WIDTH = 1e6 # Hz, half-width of the sweep window
REDIS_PORT = 6380 # <-- change to YOUR reserved port
3. Build the orchestrator#
The CalibrationOrchestrator is the main entry point. Constructing it:
loads the
example_chipdevice template (device.yaml,cluster.yaml,spi.yamlunderdevice_templates/example_chip/),builds the in-memory device model of the chip,
seeds Redis with the template defaults,
creates the
calibration_outputs/<timestamp>/output folder.
The Qblox cluster itself is connected lazily, the first time a measurement session runs.
[ ]:
orchestrator = CalibrationOrchestrator(
experiment_template_name="example_chip",
data_handler=RedisDataHandler(redis_port=REDIS_PORT),
)
4. Define the experiment parameters#
ResonatorSpectroscopyParameters is a validated (Pydantic) description of the sweep. The key fields:
Field |
Meaning |
|---|---|
|
label for this node, used in filenames & the report |
|
which qubits to sweep (names must match the device config exactly, e.g. |
|
the qubit state to prepare (0 = ground state) |
|
number of frequency points per resonator |
|
half-width of the sweep window, in Hz |
There are more fields controlling the sliding-window fit and punchout — see the API reference — but the defaults are sensible for a standard sweep.
[ ]:
res_spec_params = ResonatorSpectroscopyParameters(
name="resonator_spec",
qubits=QUBITS,
readout_state=0,
samples=SAMPLES,
half_width=HALF_WIDTH,
)
5. Run the node inside a session#
A node reads the device and hardware from the active session, so it must be created and run inside a calibration_session block. SessionOptions names the run — here the outputs land under calibration_outputs/<timestamp>/run_resonator_spectroscopy/.
We pass qoi_cls=ResonatorSpectroscopyQOI so the fit results (state-0 quantities of interest) are persisted back to the device model and Redis.
[ ]:
with orchestrator.calibration_session(
SessionOptions(session_name="resonator_spectroscopy")
):
orchestrator.run_node(
ResonatorSpectroscopy(
parameters=res_spec_params,
qoi_cls=ResonatorSpectroscopyQOI,
)
)
6. Inspect the results#
When the session closes, the run folder contains one HDF5 dataset, the session_timeline.json log, a PDF report and an interactive HTML report. The fitted quantities (resonator frequency, \(Q_l\), \(Q_e\), \(\theta\)) are written into the device model and stored in Redis as trusted values, ready for the next calibration step.
Browse everything — including on-demand replots of each resonator dip — with the bragi_quantum.gui server:
[ ]:
# In a terminal (not in the notebook):
# qc2-show calibration_outputs/
#
# Opens the archive browser at http://127.0.0.1:9999 with a live sidebar of
# every run, per-qubit plots, and the whole-pipeline view.
Next steps#
Chain several nodes in one session (resonator spectroscopy → qubit spectroscopy → Rabi) to build a bring-up pipeline — see Running Experiments.
Read Core Concepts to understand the Node / Measurement / Analysis / QOI anatomy behind every experiment.
Explore the full node catalog.